Chest Style Ultra Low Freezers -40°C To -85°C, 2 cu.ft., Manual Defrost Overview

The Chest Style Ultra Low Freezers -40°C To -85°C, 2 cu.ft., Manual Defrost (ULF-001C) from Yamato Scientific is a cascade-refrigeration laboratory freezer built for sample storage down to -85°C. This 2 cu.ft. chest-style unit holds up to 6 racks inside a 56 L chamber. LabSupplies.com supplies it as an authorized dealer.

Construction centers on a 14-gauge zinc-coated galvanized steel chamber paired with a 16-gauge steel exterior finished in powder-coated cool gray. Two hermetic compressors drive the cascade refrigeration system, while an air-cooled condenser with two heavy-duty fans removes heat without liquid coolant. CFC-free R-508B and HCFC-free R-404A refrigerants carry the thermal load.

Electrical options span 115V, 208V, and 230V at 60 Hz single-phase, plus a 220V 50/60 Hz configuration. Exterior dimensions measure 25.25" × 25.75" × 40", and the freezer ships at 335 lbs. Manufactured in the USA, this unit carries UL Listed certification.

Key Features

  • Digital temperature control: microprocessor controller with a dual display showing set point and chamber temperature, a scan button for setting, and tamper-proof lockout.
  • Cascade refrigeration: two hermetic compressors with an air-cooled, dual-fan condenser deliver sustained pull-down with no liquid coolant required.
  • Battery-backed alarm system: audible and visual signals on mechanical or electrical failure, over- and under-temperature settings, silencing switch, and battery test switch.
  • Remote alarm capability: a relay is provided for remote alarm hook-up, with remote alarm contacts standard.
  • Power failure recovery: an automatic timer system restarts the unit after a power interruption.
  • 1/2" access port: standard port supports probes and supplementary monitoring.

How to Choose the Right Ultra-Low Freezer

Selecting an ultra-low freezer comes down to matching storage demand, target temperature, and site infrastructure to the unit's design. Weigh these axes before committing.

  • Capacity: the 2 cu.ft. / 56 L chamber holds up to 6 racks, suiting focused sample sets rather than high-throughput biorepositories.
  • Temperature range: a -40°C to -85°C span covers most biological and reagent storage needs that demand sub-zero stability.
  • Control type: a digital dual-display controller with tamper-proof lockout protects set points in shared lab spaces.
  • Voltage and site power: confirm whether your facility provides 115V, 208V, 230V at 60 Hz, or 220V at 50/60 Hz before ordering.
  • Footprint: verify the 25.25" × 25.75" × 40" exterior fits the bench or floor space allotted.
  • Safety and continuity: battery-backed alarms, remote alarm relay, and automatic power-failure recovery guard samples against unattended failures.

Specifications

Specification Detail
Capacity 2 cu.ft.
Temperature Range -40°C to -85°C
Defrost Manual
Interior Capacity 56 L / 2 cu.ft.
Rack Capacity 6 racks
Interior Dimensions 17" × 14" × 14" (43 × 35 × 35 cm)
Exterior Dimensions 25.25" × 25.75" × 40" (64 × 65 × 101 cm)
Temperature Control System Digital control with dual display (set point and chamber temperature); scan button for temperature setting; tamper-proof lockout
Refrigeration System Cascade type — two hermetic compressors
Refrigerant CFC-free R-508B and R-404A (CFC- & HCFC-free)
Condenser Air-cooled with two heavy-duty fans; no liquid coolant required
Alarm System Battery-operated alarm emits an audible and visual signal on mechanical or electrical failure; over- and under-temperature settings, alarm silencing switch, and battery test switch; relay provided for remote alarm hook-up
Power Failure Recovery Automatic timer system restarts the unit after a power failure
Electrical Requirements 115V / 208V / 230V, 60 Hz, 1-phase; 220V, 50/60 Hz, 1-phase
Optional Accessories Chart recorder; CO₂ and LN₂ back-up system; racks; cold safety gloves
Material Chamber is 14-ga. Zinc coated galvanized steel. Exterior is 16-ga. steel. Power coated cool gray
Country of Origin USA
GTIN (Barcode) 797059929832
Shipping Weight 335 lbs
Dimensions (L × W × H) 31 × 31 × 48 in
Component Size 2 CF
Other Certification UL Listed

Available Sizes and Configurations

SKU Configuration Pack
ULF-001C EA EA

Documentation

More in This Series

Why Buy From LabSupplies.com

LabSupplies.com is an authorized Yamato Scientific dealer, so every ULF-001C ships as a genuine factory unit backed by the full manufacturer warranty. Order directly through secure checkout and lean on our lab-equipment specialists for spec confirmation.

  • Authorized Yamato Scientific dealer: genuine unit with full manufacturer warranty.
  • In stock: ready to ship from USA inventory.
  • Specialist support: dedicated lab-equipment team for technical and spec questions.

Frequently Asked Questions

What temperature range does the ULF-001C reach?

The ULF-001C operates from -40°C to -85°C. A cascade refrigeration system using two hermetic compressors and CFC-free R-508B and HCFC-free R-404A refrigerants sustains these ultra-low temperatures.

What is the storage capacity and chamber size?

Interior capacity is 2 cu.ft. (56 L), holding up to 6 racks. Interior dimensions measure 17" × 14" × 14" (43 × 35 × 35 cm), and exterior dimensions are 25.25" × 25.75" × 40" (64 × 65 × 101 cm).

What electrical supply does this freezer require?

The unit is available for 115V, 208V, and 230V at 60 Hz, single-phase, plus a 220V 50/60 Hz single-phase configuration. Confirm your facility's available voltage before ordering.

What materials is the cabinet built from?

The chamber is 14-gauge zinc-coated galvanized steel. Its exterior is 16-gauge steel finished in powder-coated cool gray for durability in laboratory environments.

How does the freezer protect samples during a failure?

A battery-operated alarm emits audible and visual signals on mechanical or electrical failure, with over- and under-temperature settings, a silencing switch, and a battery test switch. An automatic timer restarts the unit after a power failure, and a relay supports remote alarm hook-up.

Does this unit require manual defrosting and where is it made?

Yes, the ULF-001C is a manual-defrost chest freezer. It is manufactured in the USA and carries UL Listed certification.