The SmartDrop X NS1010 is a full-spectrum UV-Vis nano spectrophotometer that scans 200 to 800 nm in 1 nm steps using a xenon flash lamp and a 2048-element CCD array. Auto-ranging 0.2 mm and 1.0 mm path lengths let it read dsDNA from 2 to 4,500 ng/µL without dilution, from 0.5 to 2.0 µL of sample. The software reports 260/280 and 260/230 ratios, includes preset BCA, Bradford and Lowry programs, and outputs up to 5 chosen wavelengths for colorimetric work. It is operated from a 7-inch touch screen.
Key features
- 200–800 nm scans in 1 nm steps; xenon lamp needs no warm-up
- Auto-ranging 0.2 mm and 1.0 mm path lengths
- dsDNA 2–4,500 ng/µL; BSA up to 135 mg/mL
- Absorbance precision 0.003 A
- Preset BCA, Bradford and Lowry protocols; 260/230 ratios
- Cuvette port for OD600 readings and built-in printer
Specifications
| Brand | Accuris |
|---|---|
| Catalog number | NS1010 / NS1010-E |
| Display | 7 in touch screen |
| Sample Volume | 0.5–2.0 µL |
| Path Length | 0.2 mm and 1.0 mm (auto-ranging) |
| Light Source | Xenon flash lamp |
| Detector | 2,048-element linear silicon CCD array |
| Wavelength Range | 200–800 nm (1 nm steps) |
| Absorbance Range | 0.04–90 A (10 mm equivalent) |
| Absorbance Precision | 0.003 A |
| Lower Limit of Detection | dsDNA: 2 ng/µL; BSA: 0.1 mg/mL |
| Upper Limit of Detection | dsDNA: 4,500 ng/µL; BSA: 135 mg/mL |
| Preset Assays | Nucleic acid, protein, BCA, Bradford, Lowry |
| Data Export | USB |
| Electrical | 24 V DC, 2 A; 25 W |
| Weight | 6.2 lb (2.8 kg) |
| Warranty | 2 years |
Ordering information
| Catalog number | Power |
|---|---|
| NS1010 | 115V, US plug |
| NS1010-E | 230V, European plug |
Documents